Viscosity Units Explained: cP, mPa·s and cSt — with Conversion Tables

Published by A&D Gulf Technical Team  · 

Centipoise and millipascal-seconds are the same number: 1 cP = 1 mPa·s, exactly. That single fact resolves most viscosity-unit confusion — a datasheet quoting 350 cP and one quoting 350 mPa·s are stating the identical viscosity. The remaining distinction that matters is between dynamic viscosity (cP, mPa·s — what most viscometers report) and kinematic viscosity (cSt — common in oils and fuels), which are related through the liquid's density.

Dynamic vs kinematic viscosity — the one distinction to get right

Dynamic (absolute) viscosity is a fluid's resistance to flow when a force is applied — units mPa·s or cP (SI: Pa·s). This is what a tuning-fork or rotational viscometer measures, and the unit on most product and QC specifications. Kinematic viscosity is dynamic viscosity divided by density — how a fluid flows under its own weight — with units of cSt (= mm²/s). This is the petroleum world's unit: lubricants, fuels and oils are graded by gravity-driven flow.

The bridge between them: kinematic (cSt) = dynamic (mPa·s) ÷ density (g/cm³). A worked example: an oil of 90 mPa·s with density 0.9 g/cm³ has kinematic viscosity 90 ÷ 0.9 = 100 cSt. For water (density ≈ 1.0 g/cm³), cSt and mPa·s are numerically almost identical.

Dynamic viscosity — exact conversions

Petroleum labs may also meet Saybolt seconds (SUS/SSU) — an older flow-time unit whose conversion to cSt is non-linear and table-based, not a multiplication.

Unit Symbol In mPa·s Notes
millipascal-second mPa·s 1 The standard laboratory unit
centipoise cP 1 Identical to mPa·s
pascal-second Pa·s 1,000 SI unit
poise P 100 CGS unit; 1 P = 100 cP

Kinematic viscosity — exact conversions

Unit Symbol In cSt Notes
centistokes cSt 1 = 1 mm²/s exactly
stokes St 100 CGS unit
square metre per second m²/s 1,000,000 SI unit

Typical viscosity values — approximate, at about 20 °C

Every one of these values moves strongly with temperature — which is the next point.

Liquid Typical dynamic viscosity
Water ≈ 1 mPa·s
Milk ≈ 2 – 3 mPa·s
Olive oil ≈ 60 – 90 mPa·s
Glycerine ≈ 1,400 – 1,500 mPa·s
Honey ≈ 5,000 – 10,000 mPa·s

Still deciding? Tell us what you need to weigh and we will tell you which one you actually need.

Why a viscosity number without a temperature is meaningless

Viscosity falls as temperature rises — for many liquids by several percent per degree Celsius, and honey or glycerine change far faster than that. A QC spec of 500 mPa·s is only testable as 500 mPa·s at a stated temperature. Practical rules: always record the sample temperature with the reading · compare readings only at the same temperature · in Gulf facilities, remember a sample carried from a hot warehouse into an air-conditioned lab is changing viscosity as it sits on the bench.

Newtonian vs non-Newtonian: when one number isn't enough

Water and simple oils are Newtonian — viscosity is constant regardless of how fast you stir. Many real products are non-Newtonian: ketchup, mayonnaise, paints, gels, creams and slurries thin or thicken as shear increases, so their viscosity depends on how vigorously they are being worked. For these, a single reading only makes sense at a stated shear condition — and comparing instruments that shear the sample differently will legitimately give different numbers. This is why formulation labs use a rheometer rather than a single-point viscometer: A&D's RV-10000A vibro rheometer measures 0.3 – 25,000 mPa·s continuously across shear rates with real-time graphing, for exactly these products.

How viscosity is measured — and what your instrument will read

A&D's SV series uses the sine-wave vibro tuning-fork method — two sensor plates vibrating at 30 Hz — and A&D is the only manufacturer of tuning-fork viscometers. It reads continuously in mPa·s (or cP; they are the same number). The family spans 0.3 to 100,000 mPa·s across stand-type SV-A models (the SV-1A measures samples as small as 2 mL) and the handheld SV-H line for on-site checks. How the method compares with rotational (Brookfield-type) instruments is covered in our A&D vs Brookfield comparison — the short version: continuous reading, wide range without changing spindles, and small samples.

Frequently Asked Questions

Is cP the same as mPa·s?
Yes — exactly. 1 centipoise = 1 millipascal-second. A specification in either unit is directly comparable to the other with no conversion.
How do I convert cSt to cP?
Multiply kinematic viscosity (cSt) by the liquid's density in g/cm³: cP = cSt × density. For water (density ≈ 1.0) the numbers are nearly identical; for a 0.85 g/cm³ fuel, 100 cSt ≈ 85 cP.
What is the viscosity of water?
Approximately 1 mPa·s (1 cP) at 20 °C. It falls as temperature rises — one reason water is the common reference liquid for checking a viscometer.
What is the SI unit of viscosity?
Dynamic viscosity: the pascal-second (Pa·s); everyday work uses mPa·s. Kinematic viscosity: m²/s; everyday work uses mm²/s, identical to the centistoke (cSt).
What unit does a viscometer display?
Most laboratory viscometers, including the A&D SV series, read dynamic viscosity in mPa·s or cP. Kinematic (cSt) values are obtained by dividing by the sample's density.
Why do two instruments give different readings on the same sample?
Usually temperature (readings at different temperatures are not comparable) or non-Newtonian behaviour (different instruments shear the sample differently). Same temperature and same method give comparable numbers.